Ultra Low Temperature Freezer are specially designed to store samples products
Features
l Two-compressor cascade refrigeration system specially designed for cryogenic applications.
l Wide available voltage range: 187V~242V.
l Ambient temperature :10℃~32℃.
l Accurate Temperature control technology to make the temperature maintain -40℃~-86℃and accuracy 0.1℃.
l Removable heat exchanger, almost all of the relevant components of the cooling system could be maintained and replaced.
l Realize historical temperature curve query and download
l Independent switch, standard USB port, easy to download data
l HC refrigeration system, energy saving 50%
l VIPⅡPLUS technology, special narrow design for entering elevators.
Safety
l Build-in back-up battery to display temperature and to operate audible and visual alarm system for up to 72 hours without AC power.
l Visual and audible alarm system: High/low temperature, Power failure, Abnormal voltage, Sensor error, Filter screen check, Low battery, Condenser clean, High ambient temperature, Door ajar alarm
l Power failure protection: turn on delay of the cooling system when power failure, restart delay protection.
l Keyboard lock, password protection to avoid set parameter randomly.
l Special noise reduction design.
Humanized design
l 4 casters with stopper for easy moving and fixation.
l standard 25mm test hole for easy connecting to temperature monitoring systems.
l 3 adjustable shelves for easy storing.
l Visual and audible alarm and all alarm can be through reserved for remote alarm port remote alarm
l Removable condenser front cover convenient for cleaning and maintenance.
l With the function of hand-written art-board, convenient for customers to record temporary information.
l With A4 paper slot function, convenient for customer management and equipment related information
Certificate:
l ISO14001.ISO9001.ISO13485.ISO45001
Warranty:
1 year free spare parts and remote technical support.
Dual - machine cascade system.
Removable heat exchange
10 visual and audible alarms
Performance | Refrigeration System | ||
Temperature Range | -40~-86°C | Compressor/Brand | 2/SECOP |
Ambient Temperature | +10~+32°C | Refrigerant | HC |
Controller | Microprocessor | Cooling Type | Direct cooling |
Display | LCD Display | Defrost Type | Manual |
Electrical Data | Dimensions | ||
Power Supply(V/Hz) | 220/50~60; 115/60 | Capacity | 408L |
Rated Power(W) | 845W | Interior Size(W*D*H) | 891×992×1945.5 |
Electrical Current(A) | 7 | Exterior Size(W*D*H) | 606×575×1180 |
Power Consumption | 11.6kWh/24h | Package Size(W*D*H) | 954×955×2094 |
Power failure backup system | 72H | N.t./G.t.(approx) | 292/317kg |
Loading Qty(20'/40'/40'H) | 12/24/24 | ||
Alarm | Accessories | ||
High/Low Temperature | Y | Caster | 4 |
Sensor Error | Y | USB port | Y |
Power Failure | Y | Remote alarm port | Y |
Door ajar | Y | RS232/485 port | / |
High Ambient temp. | Y | Inner Door/shelf | 2/3 |
Condenser clean | Y | Test hole | 1/25MM |
Low battery | Y | chart recorder | Optional |
Filter screen check | Y | CO2/LN2 back up system | Optional |
Abnormal voltage | Y | Cold storage box loading QTY | 28800 |
Contact: Lisy
Phone: 001-719-3091089
E-mail: info@uslab.us